The present paper is concerned with the fast evaluation of the flow injection analysis (FIA) signals and the automatic correction of the analytical values interfered by systematic and stochastic disturbances. With the application of the extended Kalman filter, the highest amount of information for the data evaluation of analytical signals can be extracted from FIA peaks. The concentration of the analyte and the offset of the baseline are estimated as time-variable parameters by filtering. The results of the application to simulated and real FIA data show that the parameters corresponded to the evaluation of the analytical data, but is already available before the maximum of the FIA peak, i.e. the evaluation of the data with the Kalman filter can be done during the running peak. Good estimates of the measured values are already obtained short after start of the peak. For this reason, the measuring dead-time of the FIA system can be reduced by the use of this method. The evaluation of FIA signals disturbed often by fluctuation of the baseline and other noise can be corrected by estimation of the offset of FIA peaks and smoothing of real FIA signals. Therefore, the application of the extended Kalman filter can also improve conventional evaluation methods. Copyright (C) 1998 Elsevier Science B.V.

Organisationseinheit(en)
Institut für Technische Chemie
Typ
Artikel
Journal
Journal of biotechnology
Band
62
Seiten
11-28
Anzahl der Seiten
18
ISSN
0168-1656
Publikationsdatum
11.06.1998
Publikationsstatus
Veröffentlicht
Peer-reviewed
Ja
ASJC Scopus Sachgebiete
Biotechnologie, Bioengineering, Angewandte Mikrobiologie und Biotechnologie
Elektronische Version(en)
https://doi.org/10.1016/S0168-1656(98)00039-X (Zugang: Geschlossen)