Influence of Different Surface Machining Treatments of Magnesium-based Resorbable Implants on the Degradation Behavior in Rabbits

verfasst von
Nina Von Der Höh, Dirk Bormann, Arne Lucas, Berend Denkena, Christian Hackenbroich, Andrea Meyer-Lindenberg
Abstract

The degradation behavior of magnesium-based implants with 0.8wt% calcium and different surface conditions for 3 and 6 months was investigated. The study used MgCa0.8 implants of 3mm × 5mm cylinders receiving different surface machining treatments. The sand-blasted implants were produced from smooth implants using a downstream irradiation process. Twenty-four female, adult New Zealand white rabbits with an average weight of 3.5 kg were used for the study and kept in standardized cages with free exercise daily. The degradation of MgCa-implants with various surface modifications was examined during the study. It was observed that the hole-shaped degradation started at the thread flank affected the core particles of the screw-shaped implants. The study also found that smooth implants showed the best integration into the bone compared to sand-blasted and threaded cylinders.

Organisationseinheit(en)
Institut für Fertigungstechnik und Werkzeugmaschinen
Externe Organisation(en)
Stiftung Tierärztliche Hochschule Hannover
Typ
Artikel
Journal
Advanced Engineering Materials
Band
11
Seiten
B47-B54
Anzahl der Seiten
8
ISSN
1438-1656
Publikationsdatum
05.2009
Publikationsstatus
Veröffentlicht
Peer-reviewed
Ja
ASJC Scopus Sachgebiete
Werkstoffwissenschaften (insg.), Physik der kondensierten Materie
Elektronische Version(en)
https://doi.org/10.1002/adem.200800273 (Zugang: Geschlossen)