Immunobased elution assay for process control
- verfasst von
- Kerstin Beyer, Martin Reinecke, Wolfgang Noe, Thomas Scheper
- Abstract
A flow-injection analysis (FIA) system with an integrated immunobased elution assay was developed for the on-line determination of biotechnological products. This paper describes its application for tissue-type plasminogen activator (rt-PA) and recombinant antithrombin III (r-AT III). The short response time of the immunochemical system allows its use for measurement and control of biotechnological production processes. Immobilized polyclonal antibodies were found to bind rt-PA in the cartridge better than monoclonal antibodies. Sepharose-4B was found to be the best support for the antibodies. The cartridge with immobilized antibodies can be prepared easily, has a high reliability and is stable for more than two weeks. The total time necessary for one assay cycle is in the range of 6-8 min, depending on the conditions. The whole procedure is controlled by a personal computer. The FIA system can be used for automatic analysis as a stand-alone version with a frequency of 7 samples per hour. The standard operation range is from 500 μg/ml down to 2 μg/ml. A comparison with the conventional microtiter ELISA assay shows that this assay can be applied for an accurate determination of undiluted cultivation samples.
- Externe Organisation(en)
-
Boehringer Ingelheim
Westfälische Wilhelms-Universität Münster (WWU)
- Typ
- Artikel
- Journal
- Analytica chimica acta
- Band
- 309
- Seiten
- 301-305
- Anzahl der Seiten
- 5
- ISSN
- 0003-2670
- Publikationsdatum
- 20.06.1995
- Publikationsstatus
- Veröffentlicht
- Peer-reviewed
- Ja
- ASJC Scopus Sachgebiete
- Analytische Chemie, Biochemie, Umweltchemie, Spektroskopie
- Elektronische Version(en)
-
https://doi.org/10.1016/0003-2670(95)00092-E (Zugang: