Optical frequency ratio of a 171Yb+ single-ion clock and a 87Sr lattice clock

verfasst von
S. Dörscher, N. Huntemann, R. Schwarz, R. Lange, E. Benkler, B. Lipphardt, U. Sterr, E. Peik, C. Lisdat
Abstract

We report direct measurements of the frequency ratio of the 642 THz 2S1/2(F = 0)-2F7/2(F = 3) electric octupole transition in 171Yb+ and the 429 THz 1S0 -3P0 transition in 87Sr. A series of 107 measurements has been performed at the Physikalisch-Technische Bundesanstalt between December 2012 and October 2019. Long-term variations of the ratio are larger than expected from the individual measurement uncertainties of few 1017. The cause of these variations remains unknown. Even taking these into account, we find a fractional uncertainty of the frequency ratio of 2.5 × 1017, which improves upon previous knowledge by one order of magnitude. The average frequency ratio is νYb+ /νSr = 1.495 991 618 544 900 537(38). This represents one of the most accurate measurements between two different atomic species to date.

Externe Organisation(en)
Physikalisch-Technische Bundesanstalt (PTB)
Typ
Artikel
Journal
METROLOGIA
Band
58
ISSN
0026-1394
Publikationsdatum
02.2021
Publikationsstatus
Veröffentlicht
Peer-reviewed
Ja
ASJC Scopus Sachgebiete
Allgemeiner Maschinenbau
Elektronische Version(en)
https://doi.org/10.1088/1681-7575/abc86f (Zugang: Offen)